H.4 Reversing Auto-Calibration
A gage that has been calibrated in the field using the auto-calibration
feature is temporarily a dedicated gage, unable to be used with
interchangeable load cells. A dedicated gage may be reset to accom-
modate interchangeable Force Cell Modules by reversing the auto-
calibration function. This is accomplished by entering the setup menu,
selecting CAL and attempting to CAL with a zero load.
1. Enter the Set-Up Menu
a) Turn off the FDV
b) Press and hold SELECT
c) Press POWER
d) Release SELECT and POWER
NOTE: An "Access to Set-up Menu" label is provided on the back of the
FDV gage for quick reference.
2. Scroll through the menu until reaching “CAL”, press SELECT button to
start the calibration procedure.
3. The display will read “null”, since no weights being hung from the gage.
4. Press SELECT to accept the zero value - a short pause follows.
5. The gage will display "LXXX", where XXX is the capacity of the load cell.
6. With a zero load value displayed, press SELECT.
7. Gage displays "uuuu" (under).
8. Press the power switch.
To confirm the successful completion of this procedure, re-enter the set-
up menu and scroll to SCAL. Press SELECT. The SCAL value should
read 2.500, the calibration value necessary for free interchange of FCM
Force Cell Modules. If SCAL displays any value other than 2.500, the
Force Display Module is still dedicated to a single Force Cell Module and
is not interchangeable. The SCAL value may only be adjusted via the
Auto-calibration feature. It is not accessible in any other way.